C A U T I O N
System Configuration
If the time/date is not correctly set, or if changed during monitoring,
malfunction may occur to NIBP measurement, periodic recording, trend,
NIBP list data, and age calculation from the birth date.
When the waveform and numeric data display for each parameter is set
to OFF, the alarm generation and table/trend input for the corresponded
parameter will be also set to OFF.
If the display of waveform / numeric data labeled as BP1 or ART is set to
OFF, the pulse rate derived from BP will not be displayed either.
When the waveform and numeric data display for SpO
2
is set to OFF, the
pulse rate derived from SpO
2
will not be displayed either.
When the waveform and numeric data display for CO
2
is set to OFF, the
respiration rate derived from CO
2
will not be displayed either.
When the waveform and numeric data display for the gas module is set
to OFF, the respiration rate measured by the gas module will not be
displayed.
If “Save Data to Table” is set OFF on the alarm setup menu, data at
alarm occurrence will not be stored on the table.
Depending on the data stored on the table (alarm data, NIBP manual
measurement data, or NIBP continuous measurement data), the total
duration of the table will differ.
Do not set the same remote control bed ID to more than one monitors of
the same floor. Otherwise, it may cause to remote control more than one
monitors at the same time.
After the remote control setup, check that the remote control unit is
properly operating.
Patient Admit / Discharge
If you start monitoring a new patient without performing a discharge
procedure for the previous patient, new data will be added to the
previous data which will result in inaccuracy.
The setup for the alarm mode and display mode remains stored even
when the power is turned off or when discharging procedure is
performed. Before monitoring, make sure the current monitoring mode is
suitable for the patient’s condition.
ST Measurement
For the lead which the electrode is detached, the reference waveform
setup cannot be performed. Check if the electrode is correctly attached,
and perform the setup again.
When the electrodes are properly attached, the reference waveform
setup will be automatically performed. To display the correct ST data, the
reference point and measurement point must be set.
CF Card
Use only the specified CF card.
Use only the CF card formatted with this device.
Restart the system after reading the setup data from the CF card.
The setup data will become effective after the system is restarted.
Reading the patient data from the CF card will erase all previous patient
data stored in the patient monitor.
Maintenance
If stains cannot be removed from the touch panel surface, wipe softly
with a dry or ethanol dampened cleaning cloth. Never use strong-acidic
cleaning solution. Neither is it recommended that mild acidic or alkaline
cleaning solution to be used.
A special coating is applied to the surface of the touch panel. Do not
wipe the surface with a cloth or gauze with coarse texture. Wipe the
surface with a soft cleaning cloth provided as optional accessory or with
an eyeglass cleaning cloth.
Clean the equipment frequently so stains can be removed easily.
To prevent injury, it is recommended to wear gloves when cleaning the
equipment.
xxi
Summary of Contents for DynaScope 7000 Series
Page 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Page 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 70: ...4 4 Blank Page ...
Page 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Page 92: ...6 18 Blank Page ...