NIBP Test
: This is the test mode for the NIBP pump. A service tools is required to
perform this test.
*On this test mode, never press the START/STOP key when the
cuff is connected to the patient.
Log 1 Log 2
: Records the time and information at power ON/OFF, abnormal system
restart, exceptional handling, etc. This information can be used to find
solution when equipment failure occurs.
EEPROM
: Checks the EEPROM data (equipment information, operating time, etc.).
Refer to the section on “EEPROM” for details.
Arrhythmia Analysis : Displays the sample data for arrhythmia analysis, and the data of the
analyzed result.
Performance
: Displays interrupted data in increments.
This information can be used to find solution when equipment failure
occurs.
Maintenance (SUB-CPU)
:
Confirmation and information of the SUB-CPU are displayed. Please refer
to the items on the Maintenance (SUB-CPU) screen for details.
Log 2 (SUB-CPU) :
Information to use when an abnormality in the SUB-CPU has been
generated and is needed.
SUB-CPU Test
: The state of the SpO
2
measurement unit is displayed (1/2).
An error report is displayed when any measurement error occurs.
The clinical data recorded on the CF card is displayed (2/2).
WDT Test (SUB-CPU) :
WDT test used for factory inspection.
Normally, do not press this key.
CF Card
: This is the same setting as the “CF Card” menu under the “Settings” menu.
Module Install
: Select this when you need to install the program in the HLX-501, HLX-561
telemeter transmission module. Please do not pull out the CF card while
installing the program. CF card with module install program is required for
each module.
RTC Setup
: This is the initial setting to be performed in the factory. Normally, do not
press this key. The time will be set to the reference clock of the monitor.
The normal time setting should be performed on the “Time/Date”
setup under the “Settings” menu.
If the RTC setup is changed, operating accumulated hours, recorded time
of trend/recall and all other time will be changed which will require to
initialize the system and EEPROM.
Use this setting only when the lithium battery, RTC or CPU board has been
replaced.
Normally, there is no need to set it again after the factory setting.
Once set, please initialize the system and the operating hours of EEPROM.
WDT Test
: WDT test is used for factory inspection. Normally, do not press this key. All
the patient data etc. that have been set up will be cleared.
7
-
4
Summary of Contents for DynaScope 7000 Series
Page 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Page 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 70: ...4 4 Blank Page ...
Page 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Page 92: ...6 18 Blank Page ...