Compliance to the Electromagnetic Emissions
The DS-7000 is tested under the following electromagnetic environment.
Emissions Test
Compliance
Electromagnetic Environment – Guidance
(
The DS-7000 is intended for use in the electromagnetic
environment specified below.)
RF Emissions
CISPR 11
Group 1
The equipment uses RF energy that is necessary for the internal
functioning of the equipment itself. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby
electronic equipment.
RF Emissions
CISPR 11
Class A
This equipment is suitable for use in all establishments other than
domestic and those directly connected to a low-voltage power
supply network which supplies buildings used for domestic
purposes.
●
Compliance to the Electromagnetic Immunity (1)
The DS-7000 is tested under the following electromagnetic environment.
Immunity Test
IEC60601-1-2
Test Level
Compliance
Level
Electromagnetic Environment – Guidance
(
The DS-7000 is intended for use in the
electromagnetic environment specified below.)
Electrostatic
Discharge
(ESD)
IEC61000-4-2
±
2, 4, 6kV
contact
±
2, 4, 8kV
air
±
2, 4, 6kV
contact
±
2, 4, 8kV
air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient / burst
IEC61000-4-4
±
2kV
for power
supply lines
±
1kV
for
input/output lines
±
2kV
for
power supply
lines
±
1kV
input/output
lines
Power supply quality should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Surge
IEC61000-4-5
±
0.5, 1kV
:
differential mode
±
0.5, 1, 2kV
:
common mode
±
0.5, 1kV
:
differential
mode
±
0.5, 1, 2kV
:
common mode
Power supply quality should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines.
IEC61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(95% dip in U
T
)
for 5sec.
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(95% dip in U
T
)
for 5sec.
Power supply quality should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
If it is required to continuously operate the
DS-7000 during power failure, it is recommended
to operate on an uninterrupted power supply.
Power
Frequency
(
50/60Hz
)
Magnetic Field
IEC61000-4-8
3A/m 3A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
Note :
U
T
is the AC mains voltage prior to application of the test level.
xxviii
Summary of Contents for DynaScope 7000 Series
Page 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Page 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 70: ...4 4 Blank Page ...
Page 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Page 92: ...6 18 Blank Page ...