Software Upgrade
This paragraph explains how to perform the upgrade.
In DS-7000, the upgrade is done via a Compact Flash card (hereafter, CF card).
The upgrade of the DS-7000 main unit and NIBP module will be performed simultaneously.
1
Turn the power OFF and insert the CF card in the card slot on the right side of the monitor.
2
Turn the power ON and the upgrade will start automatically.
The installation screen will automatically be
displayed, and the installation will begin. Wait until
the indicator (bar) has reached its MAXIMUM.
※
If the installation fails, the following factors need
to be checked:
・
The installation files were not copied in the CF
card.
・
The CF card was not formatted with a DS-7000.
・
The boot block is deleted.
After the upgrade (MAIN/SUB) of the DS-7000 main
unit is completed, the upgrade of the NIBP module
will start.
※
If the installation fails, the following factors need
to be checked:
・
The NIBP module installation file was not copied
in the CF card.
5
-
2