Daily Check List
No.
Inspected Date
Inspected by
Location
Device Type
(Main Unit)
Serial No.
Device Type
(Option Unit)
Serial No.
Device Type
(Multigas Unit)
Serial No.
Item
Details
Criteria
Judgment
Appearance
Visually check the exterior for scratches, cracks,
deformation, and rust.
No abnormality should be found.
□
OK /
□
NG
Check if the unit is installed on a level surface.
The installation area must be
level and free from vibration and
shock.
□
OK /
□
NG
Installation
Check if the unit is installed in a place not
susceptible to adverse environment.
The environmental condition (ex.
temperature, humidity) of the
installed place should be as
specified. The unit should not be
subjected to splashing water.
□
OK /
□
NG
The home display appears, and
the LED located on the right side
of the display panel lights.
□
OK /
□
NG
Turn ON the monitor, and check if it operates
normally.
The date and time should be
correct.
□
OK /
□
NG
Functions
Check the control switch and touch key
operation.
They should function properly.
Key sound and alarm sound
should generate.
□
OK /
□
NG
Cables
Visually check all cables for any damage.
No damage should be found.
□
OK /
□
NG
Periodic
Inspection
Check the date of previous periodic inspection.
Should be within 1 year.
□
OK /
□
NG
Air Filter
(DS-7000)
Check the date, which the air filter was first used
(cleaned, replaced).
Used from: Day Month Year
Should be within 3 months.
□
OK /
□
NG
Fan Filter
(Multigas Unit)
Check the date, which the fan filter was first used
(cleaned, replaced).
Used from: Day Month Year
Should be within 3 months.
□
OK /
□
NG
Water Trap
(Multigas Unit)
Check the date, which it was first used
(replaced).
Used from: Day Month Year
Should be within 6 months.
It should not be filled up with
water.
□
OK /
□
NG
CO
2
Absorber
(Multigas Unit)
Check the date, which it was first used
(replaced).
Used from: Day Month Year
Should be within 1 year from
usage (replacement).
□
OK /
□
NG
O
2
Cell
(Multigas Unit)
Check the date, which it was first used
(replaced).
Used from: Day Month Year
Should be within 1 year from
usage (replacement).
□
OK /
□
NG
Comment
7
-
34
Summary of Contents for DynaScope 7000 Series
Page 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Page 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 70: ...4 4 Blank Page ...
Page 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Page 92: ...6 18 Blank Page ...