Recorder Adjustment
1) Paper Out Sensor and Mark Sensor Adjustment
1 Take out the paper from the recorder cassette.
→
Check that “No” is displayed.
2 Set the cassette.
→
Check that “Yes” is displayed.
3 Press the Paper Out Sensor Adjustment
→
Start key.
→
Check that “Adjustment was completed.” is displayed.
Check that “No” is displayed for “Rec.Paper” under the recorder status.
4 Install the recording paper inside the recorder cassette, and set the cassette.
→
Check that “Yes” is displayed for “Rec.Paper” under the recorder status.
5 Press the Mark Sensor Adjust
→
Start key.
→
Check that “Adjustment was completed.” is displayed.
→
Check that “Yes/No” is displayed for “Mark” during paper feeding.
Judgment Criteria
Cassette Sensor
Cassette status should be correctly displayed.
Paper Sensor
Paper out status should be correctly displayed.
Paper Out Sensor
Adjustment
Should be correctly adjusted, and “Adjustment was
completed.” should be displayed.
Mark Sensor
Adjustment
Should be correctly adjusted, and “Adjustment was
completed.” should be displayed.
Mark Sensor
Mark sensor should be correctly displayed.
7
-
10
Summary of Contents for DynaScope 7000 Series
Page 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Page 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 70: ...4 4 Blank Page ...
Page 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Page 92: ...6 18 Blank Page ...