W A R N I N G
When using the airway adapter, always consider the circumference of the
intubation tube. If inappropriate airway adapter is used for a patient with
low ventilation, CO
2
may mix in to the inspired air resulting in incorrect
measurement, or apnea detection may become difficult.
When multigas unit (MGU-701/MGU-702/Poet IQ) is used, the sampling
tube may get clogged by the internal condensation.
If the upper/lower alarm limit of the parameter is set to OFF, or arrhythmia
alarm is set to OFF, alarm will not function even if the system alarm is set to
ON. Pay attention when setting them OFF.
Objective and constant arrhythmia detection is possible through the fixed
algorithm incorporated in this monitor. However, excessive waveform
morphology change, motion artifact, or the inability to determine the
waveform pattern may cause an error, or fail to make adequate detection.
Therefore, physicians should make final decisions using manual recording,
alarm recording and recall waveform for evaluation.
The HR/PR alarm will not be generated unless the numeric data box
corresponded to the selected HR/PR alarm source is displayed. When
HR/PR alarm source selection is changed, be sure to display the numeric
data box corresponded to the selected HR/PR source.
The RR and apnea alarm will not be generated unless the numeric data
box corresponded to the selected RR source is displayed. If the RR source
is changed, make sure to display the numeric data box corresponded to the
RR source.
When selecting Silence or Time Only for the night mode, pay
attention not to miss any important alarm by simultaneously monitoring the
bed on other monitors such as central monitor.
When lifting this device, hold the bottom part of the main unit and not the
display unit.
About the Air Filter for Cooling Fan (DS-7000, MGU-701, MGU-702)
When the air filter is washed with neutral detergent, dry it completely
before reattaching. If the moisture is remained on the air filter, it may
damage the equipment.
The air filter must be attached. If the equipment is used without it, it may
damage the equipment.
xvi
Summary of Contents for DynaScope 7000 Series
Page 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Page 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 70: ...4 4 Blank Page ...
Page 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Page 92: ...6 18 Blank Page ...