To Check the Periodic Replacement Period
The usage duration for each “periodic replacement part” is displayed under Parts Usage Time .
Use this information as an indication for the replacement period of each part.
1
Press the Menu
→
Maintenance
→
Parts Usage Time keys.
The usage hours, months, years, or number
of measurements will be displayed for the
respective parts.
2
Make sure after replacing the parts to press the Reset key to reset the usage duration.
●
About the Maintenance of MGU-701/MGU-702/Poet IQ
The periodic maintenance of MGU-701/MGU-702/Poet IQ should be performed as follows.
Each Patient
: There should be no damage to the accessories and cables and should
be properly connected.
Replace the sampling device and sampling tube.
Every week
: Replace the water trap as necessary.
Every 2 months : Check the accuracy of O
2
sensor using the 80% O
2
mixture calibration
gas.
Every 3 months : Clean the housing of the multigas unit as necessary.
Every year
: Perform the safety test. Verify the agent gas auto calibration, and
calibrate as necessary. Replace the O
2
cell. Check the gas flow rate and
adjust as necessary. Check the CO
2
absorber, and replace if necessary.
7
-
33