5. Re-connect the cable (Pay attention to
orientation).
6. Re-install the shield.
7. Re-install the rear panel and perform
calibration.
8. Press the Menu
→
Maintenance
→
Parts Usage Time keys and press the
Reset (password “7000”) key for “CSI O
2
Cell”. Press the Yes (Hold 3sec.) key to
reset the usage for the CSI O
2
Cell.
9. Turn the power ON, and check that the multigas unit warm-up is properly performed.