Demo Start
: This key will start the DEMO mode.
The DEMO mode key will be displayed only when the rotary switch on the
main body is set to "2".
When the demonstration is finished, please make sure to turn off the power
and return the rotary switch to "0". All the patient data etc. that have been
set up until now will be cleared.
Analog Adjust
: Analog board can be adjusted.
Special adjustment tools are necessary for the adjustment of the analog
board.
Please refer to the item of the analog adjustment screen for details.
Details of the Test Menu
This section explains about the test menu items individually.
●
Maintenance
CPU A/D
Verifies the data such as voltage and analog data read by the AD converter, and checks if they are
in the normal operating range.
If “NG” is displayed, malfunction of the circuit can be considered.
Board Information
Displays board version, software version, software serial number, and other information.
Status
Displays memory capacity.
*The memory capacity will differ depending on the board version.
DIP-SW, Rotary SW
Displays the setup status of the DIP switch and rotary switch.
For details, refer to the section on “DIP Switch Setup”, “Rotary Switch Setup”.
CF Card Power
Displays whether the CF card is installed or not and the status (normal/abnormal) of the power
voltage output from the card. If any problem occurs with the CF card, the status can be checked
on this display.
7
-
5
Summary of Contents for DynaScope 7000 Series
Page 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Page 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 70: ...4 4 Blank Page ...
Page 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Page 92: ...6 18 Blank Page ...