The “SpO
2
Unit Error” message is displayed.
Cause
: There is a problem in the communication with the SpO
2
measurement unit.
Solution : A defective cable or faulty SpO
2
unit can be considered.
The “Disconnect” message is displayed.
Cause
: When the SpO
2
relay cable is disconnected during SpO
2
monitoring, this message will
be displayed.
Solution 1 : To cease monitoring, press the
(Alarm Silence) key and silence the alarm.
Solution 2 : To continue monitoring, plug in the SpO
2
relay cable. This will clear the message and
silence the alarm.
Invasive Blood Pressure
The “BP
*
Transducer OFF” message is displayed.
Cause
: The BP transducer is not connected.
Solution
: Connect the transducer.
The “BP
*
Zero Required” message is displayed.
Cause
: The BP zero balance has not been performed since the power was turned ON.
Solution
: Open the three-way valve of the transducer to air and perform zero balance.
BP value and waveform are not displayed properly.
Cause
: Blood pressure line has not been zero balanced.
Solution
: Open the three-way valve of the transducer to air and perform zero balance.
The measured data is displayed as “
×××
”.
Cause
: The BP value is outside the measurement range.
Solution
: Perform zero balance again.
The “BP Disconnect” message is displayed.
Cause
: When the BP interface cable is disconnected during BP monitoring, this message will
be displayed.
Solution 1 : To cease monitoring, press the
(Alarm Silence) key to silence the alarm.
Solution 2 : To continue monitoring, plug in the BP interface cable. This will clear the message and
silence the alarm.
Temperature
The error margin of the measurement temperature is too large..
Cause 1 : The temperature sensor connector is unclean.
Solution
: Clean the temperature sensor connector.
The numeric data is displayed as “
×××
”.
Cause
: The temperature measurement is outside the measurement range.
Solution
: Check if the temperature probe is properly inserted/attached.
6
-
12
Summary of Contents for DynaScope 7000 Series
Page 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Page 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 70: ...4 4 Blank Page ...
Page 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Page 92: ...6 18 Blank Page ...