Remote Control Code (R.C. Code)
Displays the reception status of the remote control (CF-700), which does not depend on the
“Remote Control” settings under the “Initial Settings” menu.
If the remote control is malfunctioning although the reception status displayed on this screen is
normal, check the “Remote Control” settings.
Alarm Input
Displays the status port condition.
When an external device is connected, the port condition can be verified.
Display
Displays information of the display unit.
GAS Connect Status
The connection of the gas unit is displayed.
Communication Status
Displays the status for each communication port (including communication within the equipment).
Press the Test START/STOP key to start the communication test.
This test requires service tools.
Setup
: Communication mode setup
Status
: Communication status
Transmit : Transmitting packet quantity (byte)
Receive : Receiving packet quantity (byte)
Error
: Error packet quantity (byte)
Interrupt : Interrupt packet quantity (byte)
CF Card Test
The state of writing/reading test to the CF card is displayed.
Press the Test START/STOP key to start the test.
Page Down
The next page is for development use. Do not press this key.
If the key is pressed by mistake, immediately press the Page Up key to return to the previous
page.
*Never operate on the display for development use as it may damage the equipment.
7
-
6
Summary of Contents for DynaScope 7000 Series
Page 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Page 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 70: ...4 4 Blank Page ...
Page 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Page 92: ...6 18 Blank Page ...