Precautions about the Wireless Network System
D A N G E R
When monitoring a patient with wireless telemetry, make sure the patient
data is properly received at the central monitor. Pay special attention when
channel ID at the bedside monitor is changed.
W A R N I N G
A password can be set to access the channel ID setup menu to allow only
the telemetry channel administrator to change the channel ID.
Some wireless combinations of telemetry transmitters may generate
interference with other devices.
Before selecting the channel, verify it will not interfere with other
channels.
Make sure the telemetry manager of your system is aware of any
changes to the telemetry channels.
If transmitters are used in a neighboring medical facility, your facility and
neighboring facility must make agreements on the setting of telemetry
channels to prevent telemetry interference.
C A U T I O N
If performing telemetry transmission, configure the display so that the
numeric data corresponded to the waveform is displayed. If not, the
displayed waveform or numeric data may not be transmitted.
The users should not perform the setup of channel ID and group ID as
malfunction of the equipment may occur.
BP waveform with a scale above the programmed scale cannot be
properly transmitted. When transmitting the BP waveform, check the
displayed BP waveform scale.
On a wireless network system, O
2
, N
2
O, AGT alarm will not be
transmitted to the central monitor.
On a wireless network system, the alarm generated on the DS-7000 will
be transmitted to the central monitor with some delay. For details, refer to
the operation manual for the central monitor.
xxiv
Summary of Contents for DynaScope 7000 Series
Page 46: ...MGU 701 MGU 702 Multigas Unit Optional Weight 1 8kg 1 6 ...
Page 47: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Page 70: ...4 4 Blank Page ...
Page 71: ...Software Upgrade Chapter 5 Software Upgrade 5 2 Software Upgrade Verification 5 4 5 1 ...
Page 92: ...6 18 Blank Page ...