FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
46
of
146
NO
ODD
EVEN
:
:
:
No parity
Odd parity
Even parity
[
125 CSUM
:
NO ]
Checksum at continuous and BSI formats.
NO
YES
:
:
Checksum disable
Checksum enable
[
126 CR
:
YES ]
Carriage return at continuous formats.
NO
YES
:
:
Carriage return disable
Carriage return enable
[
127 LFEED
:
YES ]
Line feed at continuous formats.
NO
YES
:
:
Line feed disable
Line feed enable
[
128 DELAY
:
50 ]
Data output delay at Modbus and BSI formats; delay between
continuous format data.
000 …. 999 milliseconds.
[
129 ADDRESS
:
01 ]
Address of the port
00 …. 99.
00 means data format without address.
[
13- RS422
]
RS 422 SERIAL PORT
Press key or
key again to enter this menu.
Or press
key to go to the next sub-block.
[
131 FORMAT
:
NO ]
Data format of the serial port
NO
CONT
FAST
PRNT
BSI
MBHL
MBLH
:
:
:
:
:
:
:
Port is disabled
Continuous data output
High speed continuous data output
Printout
BSI format for PC, PLC interface
Modbus RTU High-Low format
Modbus RTU Low-High format
Press
key to select output format different than Flintec
continuous data format while the selection is CONT in the
information display. Available functions are seen on the weight
display after pressing
key sequentially. Press
key after
selecting the data format.
Refer to Appendix 5 for details data structure.
[
FLINTEC
]
FLINTEC
HBM
TOLEDO
SYSTEC
SMA
SARTOR
RINSTR
AVERY
BASTER
LM2
:
:
:
:
:
:
:
:
:
:
Flintec continuous format.
Commonly used by HBM, GSE, PT, Systec, Rinstrum.
Commonly used by Toledo, Mettler Toledo.
Commonly used by Systec.
Commonly used by USA producers, Cardinal, Rice Lake etc.
Commonly used by Sartorious.
Commonly used by PT, Rinstrum, HBM, GSE.
Commonly used by Avery E1205.
Commonly used by Baster.
Flintec LM2 ( BX1 par.000=6 )
[
132 BAUD
: 9600 ]
Baud rate