FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
121
of
146
10.12 CC-Link IE Field
After setting related parameters you can communicate with FT-111 via CC-Link network.
There are 2 LED’s near the CC-Link connector which are;
(A) NS/RUN LED
State
Indication
Off
No power
Operation abnormal
Green
Operation normal
Red
Fatal Event
If NS/RUN and AS/ERR turn red, this indicates a fatal event.
(B) AS/ERROR LED
State
Indication
Off
No error detected (or no power)
Red
Error
An error occurred in the device. If NS/RUN is off, the device enters state
EXCEPTION. If NS/RUN and AS/ERR turn red, this indicates a fatal event.
In the case of red LED warning, check cabling, configuration and station number. Power off the instrument and
reenergize the instrument 30 seconds later.
Data Format
Data format of weight value can be programmable for Floating point (IEEE 754) or Integer. Refer to parameter
[ 191 ].
1
(A) NS/RUN LED
2
(B) AS/ERROR LED
3
Link/Activity LED (port 1)
4
Link/Activity LED (port 2)
5
P1 interface (port 1)
6
P2 interface (port 2)