FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
143
of
146
- Change E
2
ROM ( U13 ).
E28 CLOCK ERROR
Clock error.
- Change CR2032 battery.
- Change the main board.
E32 LABEL ERROR
Checksum error of EPL printout
format.
- Press enter key and re-load the EPL
code.
- Change E
2
ROM ( U13 ).
E34 NOT LOADED
ADC output is not changed for the
span calibration.
- Recalibrate the scale.
- Change mainboard.
E35 LC CONNECTIO
The load cell output is decreased
after loading.
- Check load cell connections.
- Check test weight loading.
- Recalibrate the scale.
- Change mainboard.
E36 ADD LOAD
The load is not enough for span
calibration.
E37 UNSTABLE
The load is not stable at
calibration.
- Wait until scale become stable.
- Check grounding wiring.
- Recalibrate the scale.
- Change mainboard.
E40 NO ALIBI SD
Alibi memory SD card is not
installed.
- Disable Alibi memory if not required.
- Check Alibi SD card.
- Change mainboard.
E41 ALIBI FAULT
Alibi SD card is not supplied from
FLINTEC.
- Install Flintec Alibi SD card.
E42 NEW ALIBI
Alibi memory serial number error.
The new alibi SD card is installed.
- Format the alibi memory SD
card. Refer to parameter 816.
E43 ALIBI ERROR
Alibi memory could not be
initialized.
- Check alibi memory SD card
- Change main board.
E44 ALIBI CSUM E
Alibi CSUM error.
- Check alibi memory records.
- Change Alibi memory SD card.
- Change main board.
E47 ALIBI CSUM E
Alibi information CSUM error.
E48 ALIBI SD ERR
Wrong SD card at alibi memory.
- Order Alibi SD card.
E50 DLC CARD FAI
The DLC Board is broken or not
installed.
- Re-energize indicator.
- Change the DLC Board.
E61 FLASH ERROR
E2PROM Error.
- Change main board.
E70 MB SELECT ER
Modbus selection error
- Check data format of other interfaces.
- Other interfaces should not be
Modbus.
E75 MODBUS FAULT
The Modbus SD card is not
installed in SD2 card slot.
- Order Modbus SD card.
- Install Flintec Modbus SD card.
E80 VERIFY SCALE
Reverification the scale.
- Reverify the scale after checking the
scale hardware, load cells,
performance etc.
E81 CANNOT ADDR
DLC could not addressed.
- Check the DLC connection (RS-485 &
Power supply) hardware.
- Check the DLC and S/N.
E82 SHIFT ADJUST
Shift adjustment is not available
due to load cell coefficients are out
of limits.
- Check addressing is done correctly.
- Check test weight loading on
the correct DLC.
- Check the load cell installation and
scale installation.
E83 DLC COUNT ER
No regular response from load cell
- Reenergize the indicator.
- Check the DLC connection (RS-485 &
Power supply).