6.2 Quick Access Parameter Blocks used Frequently
The instrument has fast access feature to the frequently adjusted parameters for easy usage or service. As described in the
previous section, if you press the keys below for more than 2 seconds at main block
[1
—INTERFACE ]
, you will access to the parameter blocks fast.
Fast access key
Function
Press this key to access the fieldbus parameters, Sub-block [ 19- ].
Press this key to access the digital inputs and outputs parameters, Sub-block
[ 35- ].
Press this key to access the calibration, Sub-block [ 621].
Press this key to access the diagnostic parameters, Main-block [9--].
6.3 Exit from Programming
Exit from programming after set-up the scale and calibration as described below.
1. Press
key until [ SAVE : YES ] prompt appears on the display.
2. Press
key to save the changes into the memory or
3. Press
key to change item to [ SAVE : NO ], press
key to go out from programming without
saving the changes or
4. Press
key to change item to [ SAVE : BACK ], press
key to return menu or
5. Press
key to change item to [ SAVE : TEMP ], press
key to store the changes until the
power off the instrument.
[ WAIT ] message will be shown on the display for a little while, and the weighing indicator goes back to the weighing mode.
WARNING:
Don’t forget switch off the instrument and bring the calibration DIP switch position to the “OFF”
before using your scale in trade.