FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
127
of
146
PLC Output to FT-111 Input
Bitwise of a Dword:
Dword
(R/W)
B31 B30 B29 B28 B27 B26 B25 B24 B23 B22 B21 B20 B19 B18 B17 B16
B15 B14 B13 B12 B11 B10
B9
B8
B7
B6
B5
B4
B3
B2
B1
B0
1st
Dword
(OUTPUT)
(RWr0,
RWr1
)
Next Dword defines the usage of this Dword.
2nd
Dword
(OUTPUT)
(RWr2,
RWr3)
Out
5
Out
4
Out
3
Out
2
Out
1
Expanded Commands List
Not in use
Command List
Read Data Selection
New
CMD
PLC Output to FT-111 Input 2
nd
Dword
Bit Number 2
nd
Dword descriptions
B31… B24
Set / Reset digital outputs
B23… B16
Expanded Commands List (Refer to Table 11-3)
B15… B11
Not in use
B10 … B6
Command
List
Bin
Dec
Commands
00000
0
None command is activated
00001
1
Zero
00010
2
Tare
00011
3
Clear
00100
4
00101
5
Adjust zero calibration
Calibration
00110
6
Adjust span calibration
(1)
00111
7
Total Load Cell Capacity
(1)
eCal
Coefficients
Refer to par. [613]
01000
8
Average mV/V value
(1)
01001
9
Dead Load value
(1)
01010
10
Save the coefficients of eCal
01011
11
CN (Label number)
(1)
01100
01110
12
14
Not used
01111
15
SetPoint-1 (High)
10000
16
SetPoint-2 (High)
10001
17
SetPoint-3 (High)
10010
18
SetPoint-4 (High)
10011
19
SetPoint-5 (High)
10100
20
Not used
10101
21
SetPoint-1 Low
Only available for Threshold and
Window feature.
Refer to sub-block 36-, page 59 to
see the availabilities.
10110
22
SetPoint-2 Low
10111
23
SetPoint-3 Low