FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
59
of
146
[
36- DIG OUTPUTS
]
DIGITAL OUTPUTS
(appears if option installed)
Press key or
key again to enter this menu.
Or press
key to go to the next sub-block.
[
361 OUT 1
:
NO ]
Digital output 1
Page 75,
NO
S AIN
S IND
S ANE
:
:
:
:
Disable
Absolute Indicated weight
Indicated weight
Absolute net weight
S NET
S GRO
SPC1
SPC2
STAB
NET
ZR I
ZR G
RUN
ERRO
FBUS
:
:
:
:
:
:
:
:
:
:
:
Net weight
Gross weight
Control mode-1
Control mode-2
Weighing is stable
Net weighing
Zero range of indicated weight
Zero range of gross weight
In usage
Error
Remote Output over Fieldbus or
BSI
Press
key to select output function if you need function
different than basic setpoint output and active high level. Available
functions are seen on the weight display. Press
key after set up
the output function. Refer to Section 8.2 for details.
( only at SAIN, SIND, SANE, SNET, SGRO )
[
BAS-HI
]
BAS-HI
BAS-LO
THR-HI
THR-
LO
VIN-HI
VIN-LO
:
:
:
:
:
:
Basic setpoint output, active high
Basic setpoint output, active low
Threshold, active high
Threshold, active low
Window, active high
Window, active low
[
362 OUT 2
:
NO ]
Digital output 2
Page 75,
:
[
363 OUT 3
:
NO ]
Digital output 3
Page 75,
:
[
364 OUT 4
:
NO ]
Digital output 4
Page 75,
:
[
365 OUT 5
:
NO ]
Digital output 5
Page 75,
:
[
366 ZR [d]
:
1.0 ]
Zero range in XX.X division. Default is 1.0 d.
Page 83
For example: if the scale division is 0.5 kg and this parameter
entry is 1.0 division.
The output is activated if the indication is lower than 0.5 kg.
[
367 OUTPUTS
: IMME ]
Output changing.
IMME
STAB
:
:
Immediately change even scale is not stable.
Change if the scale is stable.