FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
101
of
146
3
9 division more than Max.
4
2% more than Max.
5
5% more than Max.
40177
R/W
2
Tare type
0
Subtractive tare
1
Additive tare
40179
R/W
2
Maximum tare
Refer to par. [526] page 62
40181
R/W
2
Secondary unit
Dec
Description
0
g
(Gram
1
kg
(Kilogram)
2
t
(Ton)
3
lb
(Libre)
4
No unit
(without unit)
5
N
(Newton)
6
kN
(Kilonewton)
7
kLb
(Kilolibre)
40183
R/W
2
Reserve
40185
R/W
2
Calibration
Commands
Decimal
Definition
0
None
188
Adjust Zero Calibration command
220
Adjust Span Calibration command
( first, load test weight value to 40187 )
23205
Apply the coefficients of eCal
( first, load 40189, 40191 and 40193 )
40187
R/W
2
Span Calibration Value
40189
R/W
2
Total Load Cell capacity for eCal
eCal
Coefficients
40191
R/W
2
Average mV/V value for eCal
40193
R/W
2
Dead load value for eCal
40195
R
2
Calibration
Process Status
& Errors
Bit
Definition
B0
Ready for calibration
B1
Zero calibration in process …
B2
Span calibration in process …
B3
Calibration Timeout
- Restart calibration
B4
ADC Error
- Re-energize the instrument
- If seen again, change the board.
B5
Instrument cannot be calibrating
- Check load cell cable
- Re-energize the instrument
B6
Instrument cannot be calibrating
- Load cell signal is very low or too high
B7
Calibration Error
- Calibration loading is not enough
- Check test weight loading
- Check load cell connections
B8
Calibration load value entry Error
- Test weight is too small. Increase the weight