[ 6-- CALB / ADJUST ] Calibration and Adjustment Block
[
6-- CALIB / ADJUST
]
SCALE ZERO AND SPAN SETTING MAIN BLOCK
Press
key sequentially to access this main block,
or press key or
key to enter configuration parameters,
or press
key to go to the next block,
or press
key to exit from programming.
[
61- CALIBRATION
]
CALIBRATION
Press key or
key again to enter this menu.
Or press
key to go to the next sub-block.
[
611 TEST WEIGHT
] M Scale calibration with test weight
[
612 LINEARIZATIO
] M Multipoint scale calibration to increase linearization
[
613 ELECTRONIC
] M eCal electronic calibration without test weight
[
62- ADJUSTMENT
]
ADJUSTMENTS
Press key or
key again to enter this menu.
Or press
key to go to the next sub-block.
[
621 ZERO ADJUSTM
] M Zero adjustment
[
622 SPAN ADJUSTM
] M Span adjustment
[
623 S UNDER LOAD
] M Span adjustment of loaded scale
Span adjustment under load
Span adjustment with temporary zeroing
[
624 GRAVITY CAL
] M Gravity acceleration of the place of the calibration.
[
625 GRAVITY USAG
] M Gravity acceleration of the place of the usage.
[
63- COEFFICIENTS
]
CALIBRATION COEFICIENTS
Press key or
key again to enter this menu.
Press
key to go to beginning of the sub-block or press
key to go to the next main block.
[
631 LOAD COEFFIC
] M The load weight used at the calibration is indicated here.
[
632 ZERO COEFFIC
] M This coefficient is determined the zero point of the scale.
[
633 GAIN COEFFIC
] M This coefficient is related with the gain factor of the scale.