FT-111(D) Panel Weighing Indicator, Technical Manual, Rev.1.0.0, May 2019
Page
94
of
146
Q
Load set points
Command
: [ADR][Q][SET No][L][SIGN][SP VALUE]
Response
: [ADR][Q][STATUS]
Example :
Command
: 000123.4
Response
: 01QA
(123.4 loaded to SP1)
01QN
(Could not loaded)
01QX
(Decimal point of SP VALUE is mismatch)
Comments
:
SP Number is 2 byte ASCII char. Use 01 for SP1, 02 for SP2 and 03 for SP3.
SP VALUE data is 8-byte ASCII char with dot and non-significant zeros on the left.
R
Read set points
Command
: [ADR][R][SP No][L]
Response
: [ADR][R][STATUS][SIGN][SP VALUE]
Example :
Command
: 01R01L
Response
: 01RA+000123.4 (SP1 is 123.4)
01RN
(Could not loaded)
Comments
:
SP No is 2 byte ASCII char. Use 01 for SP1, 02 for SP2 and 03 for SP3.
SP VALUE data is 8-byte ASCII char with dot and non-significant zeros on the left.
S
Read Status
Command
: [ADR][S]
Response
: [ADR][S][STATUS-1][STATUS-2][STATUS-3]
Example :
Command
: 01S
Response
: 01SSGI
(Stable, Gross, In Range)
01SDGL
(Dynamic, Gross, Low voltage error)
Comments
:
The response includes 3 status information.
STATUS-1 can be Stable or Dynamic.
STATUS-2 can be Gross or Net.
STATUS-3
can be ‘In range’, ‘Out of range’, ‘+ Over’, ‘- Under’, ‘Low voltage’, ‘High voltage’ or ‘Errors’.
T
Tare
Command
: [ADR][T]
Response
: [ADR][T][A]
(Taring is done successfully and scale is in net)
[ADR][T][N]
(Taring could not executed)
[ADR][T][X]
(Taring is disabled)
Comments
:
The tare value is overwritten by the new tare weight value.
Status must be stable in 2 seconds time out delay. If so, Ack is send.
If it cannot be stable in time out delay, Nack is send.
U
Read digital inputs
Command
: [ADR][U]
Response
: [ADR][U][STATUS][Inputs]
Example :
Command
: 01U
Response
: 01UA03
(Input 2 and Input 1 are active)
01UA96
(Input 8,5,3,2 are active)
01UAFF
(All 8 inputs are active)
01UN
(Could not read inputs)
Comments
:
Data length change according to number of digital inputs.
Inputs are implemented to ASCII char of 4-bit
. ‘1111’ inputs are implemented to char ‘F’.
INPUTS
IN
-8
IN
-7
IN
-6
IN
-5
IN
-4
IN
-3
IN
-2
IN
-1
Bit wise
1
0
0
1
0
1
1
0
ASCII
9
6