Net weighing in the container
Taring is used to weight material in the container.
1. Place the empty container on to the scale and press
key.
2. The zero display and the NET symbol appear.
3. Check
sign on the display. If it doesn´t appear, press the
key once more for correct taring.
4. Add the material into the container and follow its weight in net.
Clearing the tare
Press
key to clear the tare. The NET symbol disappears and Gross symbol appears on the display together with the
gross weight indication. The scale instability will not impact clear the tare.
Automatic taring
The scale tares automatically and NET is displayed, after placing a weight on the empty scale, if this feature is enabled.
Refer to parameter 232 at
page 55
. The weight should be heavier than the value entered to the parameter 518 for automatic
taring.
Automatic clear the tare
The tare is automatically cleared after emptying the scale, if this feature is enabled. The scale instability will not impact clear
the tare automatically.
Restore status at power on
This function saves the tare status at power off and the instrument operates in Net at power on. This feature is used for tank
and silo weighing applications.
Printing
Press
key when the item is on the pan and weight is stable to print the label. The printout format can be selected from
parameter 161. Flintec PC software is used to program EPL printout data and download in to the instrument for free
programming the data output or for label printing including barcode. Refer to
page 49