UFS8000
Modbus
89
40055
Recorder 4
%
1
A
40056
Recorder 5
%
1
A
40057
Recorder 6
%
1
A
40058-59
Transmembrane pressure
Prog.
2
M
40060-74
Reserved
M
Measurement
I
Input
R
Relay function
A
Analogue output (current)
Measurement :
De measurement value will be stored.
When the unity and number of decimals are predefined then this will be shown in this table. When this
is programmable then this can vary and the unity and number of decimals can be read in the
corresponding registers (see § 21.1.2 “Register : unities / decimals” at page. 89).
Digital inputs :
De state of the digital input function will be read bit wise :
0x8000
Function not linked to hardware in the inventory
0x4000
Function activated
0x2000
Function activated, but within delay time
0x1000
Function activated, delay time passed
0
Function not active
Relay outputs :
De state of the relay output function will be read bit wise :
When a security switch is defined then the state of this switch will also be indicated bit wise. When a
analog output is defined the value (in %) will be indicated in the LSB byte.
0x8000
Function not linked to hardware in the inventory
0x4000
Relay active position
0x1000
Function activated
0x0400
Security switch activated
0x0200
Security switch activated but within delay time
0x0100
Security switch activated, delay time passed
0x00aa
aa = The value of the analogue output will be shown in 1%. This is the
percentage between the minimal and maximal current (0-20 or 4-20mA).
Analogue outputs :
The value of the analogue output will be shown in 1%. This is the percentage between the minimal
and maximal current (0-20 or 4-20mA).
21.1.2. Register : unities / decimals
Byte1
Byte 0
40075
Flow Feed
Decimals
Unity
40076
Flow Permeate 1
Decimals
Unity
40077
Flow Permeate 2
Decimals
Unity
40078
Flow Backwash
Decimals
Unity
40079
Pressure Feed 1
Decimals
Unity
40080
Pressure Feed 2
Decimals
Unity
40081
Pressure Backwash
Decimals
Unity
40082
Pressure Air
Decimals
Unity
40083
Pressure Concentrate
Decimals
Unity
40084
Pressure Permeate
Decimals
Unity
40085
Temperature
Decimals
Unity
40086
Transmembrane pressure
Decimals
Unity
40087-96
Reserved
0x8000 Value of register not valid
Unity :
0
No unity
20
°C
30
l/h
40
kPA
90
sec
1
%
21
°F
31
m3/h
41
bar
91
min
2
mA
32
gpm
42
psi
92
hour
33
cfm
93
day
Summary of Contents for UFS8000
Page 1: ...Controller for ultra filtration plants Operating manual Software version 1 06 UFS8000 ...
Page 7: ...40 Warranty conditions 122 ...
Page 11: ...UFS8000 Picture of front side 4 2 Picture of front side 1 LCD display touch panel ...
Page 115: ...UFS8000 Overview menu 108 30 Overview menu ...
Page 117: ...UFS8000 Terminal Block 110 32 Terminal block connection 32 1 Schematics ...
Page 118: ...UFS8000 Terminal Block 111 ...