100/180MM DISTRIBUTED GRAPHIC RECORDER: USER GUIDE
Page 129
HA029324
Issue 11 Oct 14
4.3.12 TOTALISER CONFIGURATION (Cont.)
Apply
Discard
Totaliser Number
Select totaliser
number
Enable
Select Alarm number
Alarm Number
Enable
Job Number
Category
Select: Off,
Unlatched,
Latched,
Trigger
Select Job category
Value
Total of
Low cut off
High cut off
Units
Preset
Preset now
123456
Select channel etc. to be totalised
ch units
ch units
Units of channel
being totalised
Units (totaliser)
Period scaler
Unit scaler
Descriptor
Units (totaliser)
Scale Low
Scale High
Units (totaliser)
Units (totaliser)
Zone Low
Zone High
%
%
Max Decimal Digits
Colour
Scale Type
Select None, Linear or Log
None
PV format
Select Numeric or Scientific
A/B Switching
Select Job number
1) Totaliser 1
1
Off
1
No Action
Channel 1
0
999999
Units
0
1
1
Totaliser 1
0
1
1
100
4
0
Numeric
Figure 4.3.12 Totaliser confi guration menu
Period Scaler
The totaliser equation works in seconds. If the totalised channel units are other than
‘per second’ a period scaler other than the default (1) must be entered. For example,
if the input channel is in litres per hour, then the period scaler would have to be the
number of seconds in an hour (3600).
Unit Scaler
If, for example, the input channel is in litres per hour, the totalised value will be in litres,
unless the unit scaler is set to a value other than 1. If it is more convenient, the totalised
value can be in thousands of litres by setting the unit scaler to 1000. Setting the unit
scaler negative causes the totaliser to decrement rather than increment.
Scale Low/High
The ‘zero’ and full scale values for the totaliser, as traced on the screen.
If A/B switching is enabled, a second set of scale low and scale high values can be en-
tered. ‘A’ values are used during normal operation. ‘B’ values are switched to by job
action, as described in
section 4.7
.
The remaining confi guration items are identical with the relevant items in Input Channel confi guration
(
section 4.3.3
) except that Log/Linear scale type is not available. For job information, see
section 4.7.
Note: Selecting ‘High’ compression in
Archive confi guration
may cause very high totaliser values
to be displayed incorrectly by the recorder and held incorrectly in the history fi le. This problem
may be solved by moving the relevant point to a group using ‘Normal’ compression, or by re-scal-
ing the totaliser to read (for example) TeraWatt hours instead of Megawatt hours.