4-10
setting is also to be used for the following operations, it can be accepted into this program. To do
this, only the SAVE key 6.1 must be depressed uninterruptedly long enough until “Set“ is indicated
on display 3.6 for Auto Coag 1 and the signal heard continuously during storage changes into an
intermittent signal.
WARNING
The operator and/or user of the unit must ensure that programs are not changed and stored
uncontrollably. Program changes must always be documented and all users must be informed.
Program b = Bipolar
With Program “b“, bipolar instruments connected to the bipolar
output of the unit can be used for coagulation and cutting.
The parameters for bipolar cutting with a choice of coagulation
effect are set via the AUTO CUT function field. The parameters
for bipolar coagulation are set via the AUTO BIPOLAR function
field.
Settings
AUTO CUT
The HF power output for bipolar cutting can be limited in 1 watt steps
from
100 watts
to
1 watt
.
EFFECT
can be set from 1 to 4.
AUTO COAG 1
This function field is not limited and all parameters can be set.
AUTO COAG 2
This function field is switched off.
AUTO BIPOLAR
The HF power output for bipolar coagulation can be limited in 1 watt
steps from
120 watts
to
1 watt
.
With the changeover switch for the blue pedal of a dual-pedal
footswitch,
activation can be blocked using the blue footswitch pedal.
Activation of
AUTO STOP
is possible.
AUTO START
is blocked in this program.
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...