9-2
CUTTING / ENDO CUT
Automatic arc-triggered, fractionated incision
Basic setting
t
on
= 50 ms,
t
off
= 750 ms
Soft coagulation (SOFT COAG)
HF voltage waveform
unmodulated sinusoidal alternating voltage
Crest factor C, at R
L
= 500 ohms
C = 1.4 for all setttings
Rated frequency of the HF voltage
330 kHz
Peak value of the HF voltage
max. 1900 V
P
Constancy of the HF voltage
automatically controlled
HF rated power
120 watts at 125 ohms
HF power limitation (P
HFmax.
)
from 1 watt to 120 watts in 1 watt steps
Setting of the HF power limitation
via up/down keys
Display of the HF power limitation
7-segment display, 3 decimal places
Precision of the HF power limitation
+/- 1 digit or +/- 15%
Activation of the Cutting mode
by key or pedal
HF connecting sockets
2, electronically locked from one another
Forced coagulation (FORCED COAG)
HF voltage waveform
pulse-modulated alternating voltage
Crest factor C, at R
L
=500 ohms
C = 5 at 120 W
max
, C = 11 at 5 W
max
Rated of the HF voltage
1 MHz
Peak value of the HF voltage Version 1
max. 1,300 V
p
Peak value of the HF voltage Version 2
max. 2,300 V
p
Peak value of the HF voltage Version 3
max. 2,300 V
p
Peak value of the HF voltage Version 4*
max. 2,600 V
p
HF rated power
120 watts at R
L=
350 ohms
HF power limitation
from 1 watt to 120 watts in 1 watt steps
Setting of the HF power limitation
via up/down keys
Stability of the HF power
see Diagram Forc. Coag. power via R
L
Display of the HF power limitaiton
7-segment display, 3 decimal places
Precision of the HF power limitation
+/- 1 digit or +/- 15%
Activation of the Forced coagulation
by key or pedal
HF connecting sockets
2, electronically locked from one another
Spray coagulation (SPRAY COAG)
HF voltage waveform
pulse-modulated alternating voltage
Crest factor C, at R
L
=500 ohms
C = 7 at 120 W
max
, C=15 at 5 W
max
Rated frequency of the HF voltage
1 MHz
Peak value of the HF voltage
max. 4 kV
P
HF rated power
120 watts at 500 ohms
HF power limitation
from 1
watt to 120 watts in 1 watt steps
Setting of the HF power limitation
via up/down keys
Display of the HF power limitation
7-segment display, 3 decimal places
Precision of the power limitation
+/- 1 digit or +/- 15%
Activation of the Spray coagulation
by key or pedal
HF connecting sockets
2, electronically locked from one another
*) not ICC 350 T, Z
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...