9-1
9
TECHNICAL DATA, SIGNALS, DIAGRAMS
9.1
Technical data
Cutting, monopolar with automatic voltage control (AUTO CUT)
HF voltage waveform
unmodulated sinusoidal alternating voltage
Crest factor C, at R
L
= 500 ohms
C = 1.4 for all setttings
Rated frequency
330 kHz
Maximum HF voltage at load resistance RL =
∞
650 V
P
Dynamic internal impedance in rated load range
0 ohm
Constancy of the HF voltage at RL> 350 ohms
automatically controlled
Cutting quality
4 coagulation effects selectable by key
Constancy of the 4 coagulation effects
automatically controlled
HF rated power
300 watts at R
L
= 500 ohms
HF power limitation
1 to 300 watts in 1 watt steps
Setting of the HF power limitation
via up/down keys
Display of the HF power limitation
7-segment display, 3 decimal places
Precision of the HF power limitation
+/- 1 digit or +/- 15%
Activation of the Cutting mode
by key or pedal
HF connecting sockets
2, electronically locked from one another
Cutting, monopolar with automatic arc control (HIGH CUT)
HF voltage waveform
unmodulated sinusoidal alternating voltage
Crest factor C, at R
L
= 500 ohms
C = 1.4 for all setttings
Rated frequency of the HF voltage
330 kHz
Maximum HF voltage at load resistance RL =
∞
570 V
P
Constancy of the intensity of the electric arc
automatically controlled
Cutting quality
4 coagulation effects selectable by key
Constancy of the 4 coagulation effects
automatically controlled
HF rated power
300 watts at R
L
= 500 ohms
HF power limitation
1 to 300 watts in 1 watt steps
Setting of the HF power limitation
via up/down keys
Display of the HF power limitation
7-segment display, 3 decimal places
Precision of the HF power limitation
+/- 1 digit or +/- 15%
PPS (Power Peak System)
yes
Activation of the Cutting mode
by key or pedal
HF connecting sockets
2, electronically locked from one another
Cutting, monopolar with automatic voltage control (Program "b")
HF voltage waveform
unmodulated sinusoidal alternating voltage
Crest factor C, at R
L
= 500 ohms
C = 1.4 for all setttings
Rated frequency
330 kHz
Maximum HF voltage at load resistance RL =
∞
550 V
P
Dynamic internal impedance in rated load range
0 ohm
Constancy of the HF voltage at RL> 350 ohms
automatically controlled
Cutting quality
4 coagulation effects selectable by key
Constancy of the 4 coagulation effects
automatically controlled
HF rated power
120 watts at R
L
= 125 ohms
HF power limitation
1 to 100 watts in 1 watt steps
Setting of the HF power limitation
via up/down keys
Display of the HF power limitation
7-segment display, 3 decimal places
Precision of the HF power limitation
+/- 1 digit or +/- 15%
PPS (Power Peak System)
yes
Activation of the bipolar cutting mode
pedal
HF outlet socket
bipolar
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...