4-5
2
AUTO CUT function field
All parameters can be set in this function field that are relevant to
cutting:
2.1 Setting of the coagulation EFFECT when cutting
Here the required cutting quality in regard to the coagulation effect on the cutting edges
can be adjusted.
Level 1 corresponds to minimum coagulation effect.
Level 4 corresponds to maximum coagulation effect.
2.2 Setting the power limitation
The HF power output can be limited in one watt steps from 300 watts to 1 watt. If display
2.3 shows “—“, the Auto Cut mode is switched off.
2.3 Display of the set power limitation in watts
This display shows the respectively set power limitation
2.4 HIGH CUT ON/OFF
The required Cut mode “AUTO CUT“ or “HIGH CUT“ can be set here. In “AUTO CUT“
mode, the HF voltage is constant. In the “HIGH CUT“ mode, the intensity of the arc
between the active electrode and the tissue is constant.
Activation of monopolar cutting
can be done using either the yellow key on the electrode handle or the yellow pedal on the
footswitch. Activation is visually signaled by continuous illumination of the triangle symbols in
the upper part of the CUT function field and also acoustically signaled. Monopolar cutting
instruments are connected to the CUT/COAG 1 and/or CUT/COAG 2 sockets.
Activation of bipolar cutting
proceeds via the yellow pedal. To do this, program “b“ = bipolar must be selected. Activation is
visually signaled by continuous illumination of the triangle symbol in the upper part of the Auto
Cut function field as well as acoustically signaled. Bipolar instruments are connected to the
BIPOLAR socket. The maximum power setting for bipolar cutting is 100 watts. In program “b“,
the AUTO BIPOLAR function field for bipolar coagulations can also be activated via the blue
pedal of a dual-pedal footswitch or by the white pedal on a single-pedal footswitch. For bipolar
cutting, the High Cut cutting mode is not required and therefore cannot be activated.
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...