4-7
4
AUTO COAG 2 function field
All parameters can be set in this function field and
are relevant to monopolar coagulation:
4.1
Selection of the coagulation mode
By pressing this key, one of the following coagulation modes can be selected:
4.2
Soft coagulation without Auto Stop
4.2.1 Soft coagulation with Auto Stop
4.3
Forced coagulation (Check the version. See Chapter 10.12)
4.4
Spray coagulation (Also suitable for Argon Gas coagulation)
4.5
Power limitation
The HF power output can be limited in 1 watt steps from 120 watts to 1 watt. If the
display 4.6 shows “—“, the Auto Coag 2 mode is switched off.
4.6 Display of the set power limitation in max. watt
4.7 Changeover switch for the blue pedal
If the left signal 4.7.1 is illuminated in this key, the currently selected coagulation mode of the
Auto Coag 2 function field can be activated via the blue pedal of a dual-pedal footswitch.
Activation
The coagulation modes for this AUTO COAG 2 function field can be activated via the blue key
on an electrode handle or via the blue pedal of a footswitch and stopped via this key or this pedal
or automatically (Auto Stop). Activation is visually signaled by continuous illumination of the
triangle symbols in the upper part of the COAGULATION function field as well as acoustically
signaled.
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...