7-3
1.2
If the multifunctional instrument is equipped with aspiration and/or irrigation channels,
these must be connected to the appropriate aspiration and/or irrigation devices.
WARNING
If the aspiration and/or irrigation channels of a multifunctional instrument are not used during
application in insufflated bodily cavities, the terminals for these channels must be closed with
Luer-lock caps.
2.
Connect the multifunctional instrument to the ERBOTOM ICC 350 M.
The ERBOTOM ICC 350 M is equipped with an
I
nstrument
I
dentification
S
ystem
(IIS)
.
As soon as a multifunctional instrument is connected to the MIC socket (19), the
IIS
automatically checks which instrument it is and briefly displays the code number of the
respective instrument. Then the
IIS
calls up the program defined and stored for this
instrument, whereby the ERBOTOM ICC 350 M is automatically configured especially
for this instrument. If an instrument without coding is connected to the MIC socket, the
ERBOTOM ICC 350 M remains in the basic configuration of the “C” program.
The front panel settings for both the basic configuration of the “C” program and also the
instrument-specific configurations can be changed manually, although the setting
possibilities are limited specific to the instrument in consideration of the safety of the
patients and the electrical load capacity of the instruments.
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...