4-18
4.3.5 NESSY
The units in the ERBOTOM ICC model series are equipped with
a Neutral Electrode Safety System (NESSY), which monitors both
the electric connection between the unit and neutral electrode as
well as the correct application of the neutral electrode on the
patient.
When using neutral electrodes with only one contact surface, only
the electric connection between the unit and neutral electrode is
automatically monitored. If this connection is sound, the green
NESSY signal (LED) is illuminated and all operating modes can
be activated. If this connection is interrupted, the green NESSY
signal (LED) is not illuminated and the monopolar operating modes
cannot be activated. If an attempt is made in this condition to activate
a monopolar operating mode, the red NESSY signal (LED) is
illuminated and a warning signal is heard at the same time.
When using neutral electrodes with two contact surfaces, not only
the electric connection between the unit and neutral electrode is
automatically monitored, but also the application of the neutral
electrode on the patient. Here the electric conductance between
the two contact surfaces on the neutral electrode and the skin of
the patient is constantly measured automatically and compared
with the intensity of the high-frequency current that flows through
the neutral electrode.
If the intensity of the high-frequency current is greater than is permissible at the respectively
measured conductance, NESSY then produces visual and acoustic warning signals and the operator
should only continue to use the high-frequency surgery if an incision or hemostasis has absolute
priority. If the measured conductance between the neutral electrode and the patient is too small,
monopolar operating modes cannot be activated.
The diversity of neutral electrodes models for high-frequency surgery is very large. The ideal
neutral electrode, optimal for all applications, does not yet exist. To ensure the user a large
degree of freedom in the selection of the best suited neutral electrodes for his/her purposes,
NESSY is a flexible Neutral Electrode Safety System that can be adapted optimally to the neutral
electrodes selected by the user. For appropriate advice and adaptation to suit requirements, please
contact your local Erbe office (see Chapter Addresses).
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...