10-3
Instructions for custom adaptation of the maximum time limit
1.
Call up Test program 10 as follows:
2.
Press key
2.1
with the unit switched off and switch on the power.
Pr. = Program appears in Display
2.3
The program number, e.g. 1, appears in Display
3.6.
3.
By pressing key
3.5
Ý
, program no. 10 is set and
4
.
by pressing key
2.1
, this program is activated.
The red signal lamp “Time limit” is illuminated and the respective maximum time limit
appears in the displays on the function fields.
5.
By pressing the
Ý
or
ß
key beneath the respective display, the maximum time limit for
each function field can be adjusted individually from 3 to 900 seconds.
6.
By pressing key
6.1
, the set maximum time limits are stored. The maximum time limits for
all programs, selected by key
6.2
, can be set and stored independent of one another.
7.
By pressing key
2.1
, Test program 10 can be deactivated.
8.
It is possible to end the Test mode by switching off the power switch.
WARNING
For safety reasons, a change in the automatic limitation of the maximum time limit must only be
made if all users of this unit are informed properly and in good time about this change.
In addition, a change in the automatic limitation of the maximum time limit must be properly
documented.
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...