11-1
11
CLEANING AND DISINFECTION OF THE UNIT
11.1
Cleaning and disinfection of the unit
The unit housing should only be cleaned and disinfected with nonflammable and nonexplosive
agents. Make certain here that no moisture penetrates into the unit.
We recommend a spray or wipe-down disinfection. However, the information from the disinfectant
manufacturer absolutely must be observed here.
WARNING
If cleaning or disinfection of the unit with flammable or explosive agents unavoidable, this must
be completely evaporated from the unit before switching on the unit.
Use no alcohol or disinfectant products with an alcohol base. The surface coating on the front
plate may become detached.