12-3
Error list
Error
No.
Error
How to proceed?
0
No error
1
No HF output voltage
Notify the Technical Service
2
HF output voltage too high
Notify the Technical Service
3
No HF output voltage
Notify the Technical Service
4
HF output voltage too high
Notify the Technical Service
5
LF leakage current greater than 0.05 mA
Notify the Technical Service
6
Activation error
Notify the Technical Service
7
Activation error
Notify the Technical Service
9
Time limit exceeded
Heed maximum time limit
10
Erroneous setting of the AUTO CUT
function field during activation
Before activation of a cutting mode, the AUTO CUT
function field must be completely set
11
Erroneous setting of the AUTO COAG 1
or AUTO COAG 2 function field during
activation
Before activation, the AUTO COAG 1 or AUTO
COAG 2 function field must be completely set
12
Erroneous setting of the AUTO
BIPOLAR function field during
activation
Before activation, the AUTO BIPOLAR function field
must be completely set
13
The contact surface between the neutral
electrode and the patient is too small or
the neutral electrode was not connected to
the unit.
This error is recognized and reported in
the Standby mode or during activation of
the unit.
Only for NESSY Version 2
Error in the connection between the
neutral electrode and unit
Only for NESSY Version 3
Short between the two contact surfaces of
a divided neutral electrode or in the cable
to the neutral electrode
Use a sufficiently large neutral electrode and apply to
the entire surface.
Check the connection of the neutral electrode to the
unit.
Check the connection between the neutral electrode
and unit. Probably a faulty cable or faulty plug
Eliminate short or exchange cable
17
HF leakage current greater than 300 mA
Check whether the neutral electrode was correctly
applied and make certain that the patient has no
contact with electrically conductive objects.
18
HF leakage current greater than 150 mA
Check whether the neutral electrode was correctly
applied and make certain that the patient has no
contact with electrically conductive objects.
19
The contact surface between the neutral
electrode and the patient was too small
Use a sufficiently large neutral electrode and apply to
the entire surface.
20
The neutral electrode was applied in the
incorrect direction
See Chapter 4.4.5
NESSY
21
The neutral electrode was applied in the
incorrect direction
See Chapter 4.4.5
NESSY
22
When switching on the power switch, the
yellow pedal on the footswitch or the
yellow key on the electrode handle was
already pressed
Check whether there was an operating error or
whether the yellow pedal of the footswitch or the
yellow key on the electrode handle is defective
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...