7-2
7.3
Operation
Program C = MIC (Minimally Invasive Surgery)
The “C“ program is intended for the operation of multifunctional
instruments with pneumatic actuator on the MIC connecting socket.
In this program, the parameters for multifunctional instruments
can be set. Functions which are irrelevant or even impermissible
for the operation of such instruments are limited or switched off
according to the instrument.
Settings
(e.g. for instrument 1)
The limitations for other instruments are described in the instrument instruction manual.
AUTO CUT
The HF output power for cutting with the instrument can be limited in 1
watt steps from 40 watts to 1 watt.
EFFECT can be set from 1 to 3.
AUTO CUT is blocked for electrode handles with 2 keys that are
connected to the CUT/COAG 1 socket.
AUTO COAG 1
This function field is not limited and all parameters can be set.
AUTO COAG 2
The HF power output for SOFT can be limited in 1 watt steps from
120 watts to 1 watt. For FORCED from 60 watts to 1 watt. SPRAY is
blocked.
Activation of AUTO STOP during SOFT is possible.
Electrode handles with 2 keys that are connected to the CUT/COAG 2
socket cannot activate the unit.
AUTO BIPOLAR
This function field is blocked. In the display, the position of the needle
of the multifunctional instrument appears.
“-nd” means needle extended,
“nd” needle withdrawn.
Operation of multifunctional instruments with pneumatic actuator on an
Erbotom ICC 350 M
1.
Installation
1.1
Connect the ERBOTOM ICC 350 M to a compressed air source with 5 +/- 1 bar. As a
compressed air source, the central compressed air supply of the OR can be used. To connect
the ERBOTOM ICC 350 M to the standard gas supply outlet according to DIN 13 260 Part
2 in the operating room, use one of the following ERBE compressed air tubes:
ERBE compressed air tube, 5 m long, Article No. 20128 - 000
ERBE compressed air tube, 1 m long, Article No. 20128 - 001
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...