12-2
12.3
Automatic error documentation
The ICC is equipped with an automatic error recognition, error signaling and error documentation
device. Errors are recorded which are present or occur during the switching on of the power
switch, during the Self Check after switching on the power switch, during tests or during activation
of the unit and/or of accessories connected to the unit. The smaller errors recognized by the unit
are only signaled visually, larger errors are signaled visually and also acoustically. The various
errors recognized by the unit are assigned error numbers. If an error occurs, it is not only
immediately signaled visually and/or acoustically, but also the corresponding error number is
stored in the unit, where it then also remains stored once the unit has been switched off. The last
10 error numbers can be called up as follows via Test program number 2 at any time.
Calling up the stored error numbers via Test program 2:
a)
Press key 3 while the unit is switched off and switch on the power switch.
·
Pr (Program) appears in the AUTO CUT display
·
1 appears in the AUTO COAG 1 display
b)
Select program no. 2 by pressing key 8
·
2 appears in the AUTO COAG 1 display.
c)
Test program no. 2 is activated by pressing key 3.
·
Err (Error) appears in the AUTO CUT display.
·
1 (error memory location no. 1) appears in the AUTO COAG 1 display.
·
The error number stored at error memory location 1 appears in the AUTO BIPOLAR
display.
d)
In the error list, the various errors for the error numbers are described.
e)
By pressing keys 8 or 9, the 10 error memory locations available can be called up.
·
The number of the current error memory location appears in the AUTO COAG 1 display.
·
The current error number appears in the AUTO BIPOLAR display.
f)
By pressing key 7, the error numbers on all error memory locations can be deleted.
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...