12-4
23
When switching on the power switch, the
blue pedal on the footswitch or the blue
key on the electrode handle was already
pressed for CUT/COAG 1
Check whether there was an operating error or
whether the blue key on the electrode handle or the
cable to the electrode handle is defective
24
When switching on the power switch, the
blue pedal on the footswitch or the blue
key on the electrode handle was already
pressed for CUT/COAG 2
Check whether there was an operating error or
whether the blue pedal is defective
25
When switching on the power switch, the
blue pedal on the dual-pedal footswitch
or the white pedal on the single-pedal
footswitch was already pressed
Check whether there was an operating error or
whether the blue pedal of the dual-pedal footswitch or
the white pedal of the single-pedal footswitch is
defective
26
When switching on the power switch,
there was already an electrically
conductive connection between the two
poles of the bipolar instrument
Always make certain that bipolar or monopolar active
electrodes or instruments are laid down on electrically
nonconductive surfaces
27
The internal temperature of the unit was
too high. The maximum HF power was
automatically reduced.
Always set up the unit in such a way that air can reach
the housing. If this error number appears, the unit may
continue to be used if the reduced power is sufficient
for the respective operation
28 to 29
These errors concern functions in the unit
Please notify the Technical Service
30
The load resistance of the unit was too
low
Either the contact surface of the active electrode was
too large or the power limitation set too low, or there
was contact between an active electrode and a metallic
instrument, e.g. trocar sheath.
31
The HF generator of the unit was
overloaded too long. The maximum
power output was automatically reduced.
The unit can briefly supply more than 300 watts
power. More than 400 watts, averaged over 1 second,
are not permitted for reasons of safety.
32 to 35
These errors affect functions in the unit
Please notify the Technical Service
36
The Auto Start key was pressed while the
coagulation electrode was already
touching the tissue
Always first set the required activation mode and then
apply the coagulation electrode to the tissue to be
coagulated
37
This error concerns functions in the unit
Please notify the Technical Service
38
Short between the two poles of a bipolar
instrument, i.e. a bipolar forceps
Avoid shorts
40 to 49
This error concerns functions in the unit
Please notify the Technical Service
50 to 51
Error in a front panel key
Please notify the Technical Service
71 to 76
These errors concern the serial interface
on units that are equipped with a serial
interface
Please notify the Technical Service
80 to 91
These errors concern units with built-in
Neurotest
81
Technical error in the intensity setting
Please notify the Technical Service
82
Neutral electrode missing or too small
Check neutral electrode
83
Pulse current somewhat too high,
however still within the tolerance limits
84
Pulse current too high
Please notify the Technical Service
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...