9-5
9.2
Visual and acoustic signals
STATUS
visual
acoustic
HF off
Power switch on
•
Operating error on the front panel
••
••
••
HF generator on
••
••
LF leakage current greater than 0.05 mA
••
Error No.5
••
••
HF leakage current greater than 150 mA, but less than 300 mA
••
red
HF leakage current greater than 300 mA
••
red
Error No. 17
••
Output error
••
red
••
••
Max. activation time limit reached
••
red
Max. activation time limit exceeded too long
••
red
••
••
No neutral electrode connected to the unit
••
red
Error No.13
••
••
Single-surface electrode connected to the unit
•
green
Dual-surface neutral electrode connected to unit and contact
resistance between the contact surfaces less than 120 ohms
•
green
Dual-surface neutral electrode not applied to patient or contact
resistance between the contact surfaces and patient greater than 120
ohms
••
red
Error No.13
••
••
HF current relative to the contact resistance between the contact
surfaces of a dual-surface neutral electrode and patient too great
••
red
Error No.19
••
4
times
HF partial currents of the contact surfaces of a dual-surface neutral
electrode different
••
red
HF partial currents of the contact surfaces of a single-surface neutral
electrode too different
••
red
Error No.21
••
••
Interruption of the line between unit and neutral electrode or between
patient and dual-surface neutral electrode during HF activation
••
red
Error No.13
••
••
Auto Stop for Auto Coagulation
••
••
Activation error
••
••
••
Error message upon error recognition during self check
•
•
•
•
These signals are produced regardless of whether the unit is activated or not activated.
••
These signals are produced only if the unit is activated.
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...