8-3
Protocol parameters
Baud rate:
19,200
Start bit:
1
Data bit:
8
Stop bit:
1
Parity:
even, no
Software installation
Two files are provided, ICC3.EXE and EGAVGA.BGI.
EGAVGA: graphic driver.
ICC3.EXE: ICC 350 documentation program.
Program start with ICC3.exe
IMPORTANT
The program only runs reliably in DOS.
ICC 350 incision image program
The activation time for the incision image registration is 32 seconds max.. If activated for longer,
the messages “Protocol error“ and ERROR 73 appear.
Function keys in the main menu
F1
or
F8
:
Incision management.
The incisions are automatically numbered in sequence. In incision management, an incision
can be loaded using F1. The incision data and the unit settings are displayed in the main menu
for the loaded incision.
F3
:
Activation readiness.
The PC is ready to receive incision data. The ICC and the PC constantly check the incision
location. As soon as the connection is OK, the message “ICC connection OK“ appears and a
point is seen on the seven-segment display for the ICC program number.
F9
:
Incision image display.
With F9, the diagram for the displayed incision number is shown.
F10
:
Quit.
Program end.
Function keys in the incision image display
In the upper half, the program shows a sectional enlargement. The section shown is marked in
the lower half by two vertical lines. The section can be shifted using the mouse or the cursor
keys.
F1
:
Display diagram HF output voltage ON or OFF.
The voltage is displayed in effective volts.
F2
:
Display diagram HF output current ON or OFF.
The actual output current is displayed in mA.
F3
:
Display diagram resistance ON or OFF.
The resistance is limited to 30 kOhm in the diagram.
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...