4-4
4.2
Description of the controls
This symbol, in accordance with EN 60 601-1, is intended to indicate to the user
that this unit must only be used on the patient if the user is acquainted with the
operation and features of this unit.
The figures set in cursive relate to the ICC illustration for this chapter, or to the function
fields in the text.
1
Power switch
Using this power switch, the unit is switched on and off.
Each time after being switched on, the unit automatically proceeds with various performance
checks. If an error in the unit or in the accessories is recognized here, a warning signal
sounds and the determined error is indicated by a corresponding error number. (See Chapter
12.1, Automatic performance checks after switching on the unit). If no error is determined,
the unit is ready to operate.
If the unit was switched off for less than approx. 15 seconds, all settings for the program
used before switching off appear after the automatic performance check on the front panel,
and the unit can be immediately reactivated. This is advantageous if, for example, the
power supply briefly fails.
If the unit was switched off for longer than approx. 15 seconds, the basic setting of the
program used before switching off appears after the automatic performance check on the
front panel, whereby all relevant visual displays continue to blink and the unit cannot be
activated until any key on the front panel is briefly pressed as confirmation that this program
should be used. Then the relevant displays are continously illuminated and the unit can be
activated using the available settings. These settings can be changed or adapted to the
current requirements at any time. However, other programs can be selected as well.
2 - 5 Function fields
The AUTO CUT, AUTO COAG 1, AUTO COAG 2 and AUTO BIPOLAR function fields
can be adjusted separately from one another, although not activated simultaneously for
reasons of safety.
WARNING
Function fields that are not used may be switched off completely to prevent unintentional
activation. To do this the power limitation must be turned down so far in the corresponding
function field until a beep is heard and “—“ appears on the digital display. The corresponding
function field cannot be activated in this condition.
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...