During every activation of the NEUROTEST device, the pulse generator is monitored for
unit-related output error. Here, every pulse is automatically monitored. If the amplitude of
a pulse is greater or less than the currently set amplitude, the unit produces warning signals.
If the amplitude of a pulse exceeds a maximum value, the output circuit is automatically
switched off immediately. If the amplitude was so high that the electronic fuse was triggered,
this error is indicated by error no. 90.
c)
Every time after switching on the NEUROTEST device, the function of the electronic fuse
hardware is automatically tested. If an error in the electronic fuse hardware is determined,
this is indicated for a few seconds by error no. 91.
d)
The maximum activation time for the NEUROTEST device is automatically limited for
reasons of safety. After 80 seconds of uninterrupted activity, the “time limit” visual signal
is illuminated and after a further 10 seconds, the Neurotest device is automatically
deactivated. However, the Neurotest device can be immediately reactivated after this.
6.4
Technical data
NEUROTEST device in the ERBOTOM ICC 350 T
Pulse waveform
square
Pulse duration
50 ms
Pause length
550 ms
Pulse amplitude
Step 1 = 05 mA
Step 2 = 10 mA
Step 3 = 15 mA
Step 4 = 20 mA
Step 5 = 25 mA
Constancy of the amplitude
automatically controlled
Acoustic signals
synchronous with the current pulses from at least
3 mA
Prevention of unit-related output error
yes
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...