4-14
14
Terminal for potential equalization
For this, see Chapter 10.4, INSTALLATION.
15
Volume of the acoustic signal
The volume of the acoustic signals can be adjusted with this knob.
This does not apply to warning signals, which must always be sufficiently loud.
WARNING
An important purpose of this acoustic signal is to protect the patients and personnel from burns
due to unintentional activation of the high-frequency generator (for more information, see Chapter
3.1.2, Unintentional activation of a high-frequency generator).
16
Loudspeaker for acoustic signals
Always set up the instrument in such a way that acoustic signals are easily heard from this
speaker. The volume of the acoustic signals is set at the factory.
17
Power connection
This high-frequency surgical unit must only be connected via the power cord supplied by the
unit manufacturer or one of these of equal quality, which bears the national test symbol, to
correctly installed hospital grade power sockets. Here, for reasons of safety, no multiple sockets
or extension cords must be used if possible. If their use is unavoidable, they must be equipped
with a correctly functioning grounded connector.
18
Power fuses
The unit is secured with two fuses. If these fuses fail, an authorized technician should inspect the
unit for possible errors before putting back into operation.
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...