8-1
8
ERBOTOM ICC 350 M-DOKU
8.1
Intended use
As a supplement to the ICC 350, the ERBOTOM ICC 350 M-DOKU is intended for the operation
of multifunctional instruments with pneumatic actuator and MIN instruments. For this, it is
equipped with the programs C = MIC (minimally invasive surgery) and n = MIN (minimally
invasive neurosurgery) and additional controls.
Furthermore, this high-frequency surgical unit can be connected to a PC. With the PC and the
DOKU program, the parameters of the unit can be recorded via accessories during activation.
Regarding the functions of the MIC / MIN, please read Chapter 7.
8.2
DOKU system
With a PC and the DOKU program, the parameters of the unit (voltage, current, resistance,
power, cosine Phi, spark intensity and energy over time) can be recorded during activation via
the accessories (optical waveguide cable, optical waveguide box). There is no electrical connection
between the computer and the ICC 350 M-DOKU, since communication is via optical waveguide.
Installation
1.
Connect the optical waveguide box to the COM1 interface of the computer using the 9-pin
Sub-D cable.
2.
Switch on the computer and the ICC 350.
3.
Connect the optical waveguide cable to the optical waveguide box. On the other end of the
optical waveguide cable, one of the two connecting plugs lights up in red.
4.
Connect the illuminated connecting plug with the non-illuminated optical waveguide socket
of the ICC 350.
5.
Connect the other connecting plug from the optical waveguide cable to the illuminated
connecting socket of the ICC 350.
Set-up instructions
The PC or the laptop with the optical waveguide box must be set up outside the prescribed
patient surroundings. This means further than 1.5 m from the patient table. The distance to other
units in the OR and the ICC 350 M-DOKU must also be greater than 1.5 m. The computer, the
optical waveguide box and the keyboard must not be able to be touched from the patient
surroundings.
WARNING
The computer must never be placed on or under the ICC 350. The minimum distance of 1.5 m to
the patient, the units, the OR and the ICC 350 absolutely must be maintained.
Summary of Contents for ERBOTOM ICC 350 M
Page 1: ...ERBOTOM ICC 350 Instruction manual 08 00 ERBE...
Page 2: ......
Page 8: ...1 2...
Page 10: ...2 2...
Page 18: ...3 8...
Page 19: ...7 8 9 10 International Standard...
Page 20: ......
Page 42: ...4 22...
Page 78: ...9 20 HIGH CUT Effect 1 2 3 4 Power output as a function of the power limitation RL 500 ohms...
Page 81: ...9 23 SPRAY COAGULATION Power output as a function of the power limitation RL 500 ohms...
Page 88: ...11 2...
Page 94: ...12 6...
Page 96: ...13 2...
Page 98: ...14 2...
Page 100: ...15 2...