3.4
Water Level Sensor
The water level sensor (WLS) probe, with integrated temperature elements, is attached at the lower end
of the flexible protection tube. A weight is attached to stabilize the tube as illustrated in
. At
the upper part of the sensor probe, nuts are placed at the middle of the threaded section, 350 mm below
the top of the probe. This is intended as a starting point for adjusting the vertical position of the probe.
Figure 3-4. Water Level Sensor with integrated temperature sensors
As an option, the tube may be stabilized by putting a concentric weight above the WLS probe, instead of
at the end, in order to ensure that measurements are performed as close to the tank bottom as possible.
Also, the eyebolt at the end of the tube can be removed.
See also
“Water Level Sensor calibration” on page 53
“Water Level Sensor measuring range” on
for more information on how to calibrate and configure the water level sensor.
Upper Sensor Limit (100%)
Lower Sensor Limit (0%)
WLS Probe Active Length
Standard: 500 mm (19.7 in.)
Option: 1000 mm (39.4 in.)
Recommended minimum distance: 1 m ( 3.3 ft)
Weight
3
50 mm
(1
3.8
in.)
Nuts to adjust
vertical position of
the sensor probe
Summary of Contents for Rosemount 2240S
Page 2: ......
Page 10: ...viii Reference Manual 00809 0100 2240 Rev DA Title Page October 2017 Title Page ...
Page 16: ...6 Reference Manual 00809 0100 2240 Rev DA Introduction October 2017 Introduction ...
Page 28: ...18 Reference Manual 00809 0100 2240 Rev DA Overview October 2017 Overview ...
Page 192: ...182 Reference Manual 00809 0100 2240 Rev DA Index October 2017 Index ...
Page 193: ......