98
Reference Manual
00809-0100-2240, Rev DA
October 2017
4. See sections
“Water Level Sensor calibration” on page 53
“Water Level Sensor measuring range”
for details on how to calibrate and configure the Water Level Sensor. These sections
describe the tank geometry and how to calculate the Level Offset (X) parameter.
F
OUNDATION
Fieldbus parameters:
TRANSDUCER BLOCK 1100>SB_STATUS
TRANSDUCER BLOCK 1100>WLS_CALIBRATION
TRANSDUCER BLOCK 1100>SB_UPPER_DEAD_ZONE
TRANSDUCER BLOCK 1100>SB_LOWER_DEAD_ZONE
TRANSDUCER BLOCK 1100>WATER_LEVEL
5. Click the
Finish
button to finish the WLS configuration and close the Sensor Setup window.