95
Reference Manual
00809-0100-2240, Rev DA
October 2017
6. The temperature elements are numbered from the bottom of the tank and upwards. Enter the
position of each element, measured as the distance from the Zero Level (Dipping Datum Plate) to the
temperature element. If you use average temperature elements, enter the position of the terminating
level of each sensor element.
F
OUNDATION
Fieldbus parameters:
TRANSDUCER BLOCK 1300>INSERTION_DISTANCE
TRANSDUCER BLOCK 1300>SENSOR_POSITION_N
7. Click the
Finish
button.