EK280
Application Manual
150
9.7
Updates archive
Attempted or completed updates of the device software are archived in the updates archive via an interface.
The content of the update archive can be shown on the display of the device via the path
"Serv." > "Archives"
> "Update archiv"
.
i
The update archive
is not a flexible archive.
9.7.1
Archive content
Pos.
Address
Display
Description
Retrievable
1
01:01F8 GONo
Global Block number
S
2
08:0A20 ABNo
Archive Block number
A + S
3
01:0400 Time
Date and time
A + S
4
02:0197 NVers
New firmware version
A + S
5
01:0198_1 SigRL
Firmware signature part 'R', low part
A + S
6
01:0198_2 SigRH
Firmware signature part 'R', high part
A + S
7
02:0198_1 SigSL
Firmware signature part 'S', low part
A + S
8
02:0198_2 SigSH
Firmware signature part 'S', high part
A + S
9
08:0811 User
User who started the firmware update
A + S
10
08:0812 Reslt
Result entries of the software updates
A + S
11
08:0A21 Check
Check sum result ( „CRC OK“ or „Error“)
A + S
The values can be retrieved in accordance with the information in the table via D = display and/or
I = interface.
9.7.2
Memory capacity
The archive can store the last 20 updates or update attempts on the device software.
i
The updates archive is not a ring-type memory.
If the archive is full, then the archive must be deleted in order to carry out further updates or to
attempt to do so (see Chapter 9.7.3).
9.7.3
Delete updates archive
i
The update archive
is also deleted via the function "
ClrV
" (see Chapter 2.4.2
So that the updates archive is deleted from the device, open the calibration lock as described in the
EK280 operating instructions.
Move the cursor to
"Serv."
>
"Archives" > "ClrUL".
Press the ENTER button
"0" flashes.
Change the value using the arrow buttons
or
to
"1"
.
Press the ENTER button to confirm the set value. It is possible to abort the entry by pressing the
ESC button.