EK280
Application Manual
11
2.4
Special commissioning function
2.4.1
Device software and checksum
Move the cursor under tab "
Serv.
" to
"Identification" > "Volume Converter" > "Vers"
or
"Chk"
The checksum
"Chk"
can be recalculated for verification purposes by pressing the ENTER button.
2.4.2
Clear volumes and archives ("Clr.V")
i
The
"Clr.V"
functions described here delete all volume counters and all archives and logbooks.
So that
all
volume counters and all archives and logbooks are deleted from the device, open the
calibration lock as described in the EK280 operating instructions.
Move the cursor in tab
"Serv."
to
"Edit and delete” > "Clr.V".
Press the ENTER button
"0"
flashes.
Change the value using the arrow buttons
or
to
"1"
.
Press the ENTER button to confirm the set value. It is possible to abort the entry by pressing the
ESC button.
2.4.3
Clear archives ("Clr.A")
i
With the
"Clr.A"
functions described here, only the measurement period archive, the daily archive,
monthly archives 1 and 2 and the four user archives can be deleted.
So that the measuring period archive, the daily archive, monthly archives 1 and 2 and the four user
archives are deleted from the device, go to and open the administrator or calibration lock as
described in the EK280 operating instructions.
Move the cursor in tab
"Serv."
to
"Edit and delete > "Clr.A"
.
Press the ENTER button
One digit of the value flashes.
A safety function is provided to prevent accidental deleting of meter counts and archives. The serial
number of the unit must be entered to activate the safety function. For input, move the cursor with
the arrow keys
or
to the digits and change these with the arrow keys
or
.
i
The serial number of the device is printed on the identification label or it can be fined in the
"Serv."
register under
"Identification" > "Volume Converter" > "SNo"
.
Press the ENTER button to confirm the value setting. Input can be canceled using the ESC key.
2.4.4
Parameterize the volume at measurement conditions "Vm"
i
A one-off setting of the volume counter is possible from software version 2.20 while the
administrator lock is open.
The counter of the EK280 can be set once in a similar way to that of the gas counter while the administrator
lock is open for recording the volume in practice. When the calibration lock is open it is always possible to
set the volume and this can also be done directly via the operational volume (tab
"Serv."
to
"Volume" >
"Actual volume" > "Vm"
.