EK280
Application Manual
52
7.1.3.3 Retrieving data with enSuite
a) Opening the test time window for the following test retrievals with enSuite
i
The test time window only needs to be open if none of the time windows 1 to 4 are open.
If the communication link to the EK280 via the optical interface is not set up, it must be created. (see
chapter 3.1 "Setting up the communication link to the EK280 via optical interface", page 15)
Select "
Navigation
" on the left-hand side of the enSuite window and then "
Parameterize online
" at
the bottom of the navigation window.
In the top of the device window open the path "
Parameters
" > "
Interfaces
" by clicking on the relevant
plus signs, and then select "
internal modem
" (not the plus sign before).
Click on the "
0
" value next to "
Test call acceptance window
" in the right-hand side of the window and
enter a duration in minutes (i.e. 10), for the test time window for the test retrieval to remain open. For
a battery-powered EK280 please do not enter an excessively long duration, as this will adversely
affect the battery life.
Press the "
Transfer changes to device
" icon at the top of the enSuite window.
The EK280 is now available via the modem for the duration set.
In enSuite press the "
Disconnect
" or "
Disconnect from all devices
" icon to terminate the optical
interface communication link between enSuite and the EK280.
b) Test call up through a GSM modem connection
Set up the modem communication link between enSuite and EK280. (see chapter 3.2 "Setting up the
communication link to EK280 via a modem", page 15)
Once the communication link is set up, you can carry out checks, for instance view certain
parameters. To do this press "
Navigation
" on the left-hand side of the enSuite window and then
"
Parameterize online
" in the bottom of the navigation window, then click on the plus sign next to
"
Parameters
" and then its subsections.
In enSuite press the "
Disconnect
" symbol or "
Disconnect from all devices
" to end the test retrieval.
c) Test call up through the TCP/IP network.
Set up the network communication link between enSuite and the EK280. (see chapter 3.3 "Setting
up the communication link to EK280 via a network (TCP/IP)", page 16)
Once the communication link is set up, you can carry out checks, for instance view certain
parameters. To do this press "
Navigation
" on the left-hand side of the enSuite window and then
"
Parameterize online
" in the bottom of the navigation window, then click on the plus sign next to
"
Parameters
" and then its subsections.
In enSuite press the "
Disconnect
" symbol or "
Disconnect from all devices
" to end the test retrieval.